SCPI
COMMAND SUMMARY
frequency limits and multiplier will be altered accordingly. However,
the leveling point is not changed (see POWer:ALC:SOURce).
l
SYSTem:
PRESet
[
:
EXECut
el
Sets the instrument to its “local operation” state. This is the same
as pressing the front panel “green” key. There is no corresponding
query.
l
SYSTem:PRESet:SAVE
Saves the present state so it can be used whenever the command
SYSTem:PRESet
[:EXECutel
is executed or the front panel “green”
key is pressed.
l SYSTem:PRESet:TYPE
FACTorylUSER
a
SYSTem:PRESet:TYPE?
Sets and queries the type of preset to execute when the
SYSTem:PRESet
[:EXECute] command is given. Factory preset
defaults all values to factory specified values. User defined preset
defaults all values to a specified state of the instrument that you have
saved with SYSTem:PRESet
:SAVE.
l SYSTem:SECurity:COUNt
Cnum>lMINimumlMAXimum
l SYSTem:SECurity:COUNt?
[MINimumIMAXimum]
Sets and queries the number of times to clear memory when the value
of SYSTem:SECurity[:STATe] changes from ON to OFF.
*RST value is 1.
l
SYSTem:SECurity[:STATe]
ONlOFFlllO
l
SYSTem:
SECurity
[
:
STATe]
?
Controls the security mode of the instrument. When in secure
mode, any display annunciators that have been disabled cannot be
re-enabled without destroying certain stored information. This value
is not affected by *RST. When you change the value from ON to
OFF, everything except calibration data is initialized or destroyed.
In particular, data in instrument state and all save/recall registers
are destroyed.
l
SYSTem:VERSion?
This query returns a formatted numeric value corresponding to
’
the SCPI version number to which the instrument complies. The
response has the form YYYY.V where the Ys represent the year
version (i.e. 1990)
and the V represents an approved revision number
for that year. This is a query only and therefore does not have an
associated
*RST
state.
l
TRIGger
[
:
IMMediat
e]
Causes the trigger event to occur regardless of other settings in the