SCPI
COMMAND SUMMARY
subsystem. This event does not affect any other settings in this
subsystem.
This command has no effect unless the instrument is in the wait for
trig state. If the swept CW generator is in the wait for trig state,
it performs its trigger action. This is an event and has no “RST
condition.
l
TRIGger:ODELay
<num>[time
suffix]
lMAXimumlMINimum
l
TRIGger:ODELay?
[MAXimumlMINimum]
Sets
and queries the trigger output delay, the time between when
the source is settled, (when Bit 1 of the Standard Operation Status
Register makes a negative transition), and the trigger out signal is
sent.
l
TRIGger:SOURce
IMMediatelBUSlEXTernal
l
TRIGger:SOURce?
Sets and queries the source of the trigger event.
l
TSWeep
This is a convenience command that does the equivalent of
ABORt;INITiate[:IMMediate].
l
UNIT:POWer
[lvlsuffix]
l
UNIT:POWer?
Sets and queries the default power subsystem units.
*RST value
is DBM.
Operating and Programming Reference S-47