3-14-1- Emptying the solid waste
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
Waste which contains, or has been in contact with, biological specimens must be
considered to pose a potential risk of infection.
Wear disposable gloves when handling waste.
The waste must be disposed of in accordance with current local regulations.
Open the compartment by pushing on the cover, then remove the waste tray.
With re-usable solid waste
Empty the content of the solid waste tray into a container reserved for disposal of biological waste.
Waste must be disposed of in accordance with current local regulations (see APPENDIX I: Waste disposal,
page A2).
Decontaminate the solid waste tray (see Section 5, 5-5-1- Decontaminating the containers of solid and
liquid wastes, page 147).
With disposable solid waste container
Remove the disposable container from the solid waste tray.
Discard the disposable solid waste container and its content into a container reserved for disposal of
biological waste.
Waste must be disposed of in accordance with current local regulations (see APPENDIX I: Waste disposal,
page A2).
Install a new disposable container into the solid waste tray.
Return the waste tray inside its compartment.
Close the cover by pushing on it until it remains locked.
Then release the solid waste alarm by clicking on .
Click on the solid waste icon.
A confirmation message displays.
Click on to confirm the emptying of the waste.
Exit the ancillaries management window.