5-4- Analyser Interventions
5-4-1- Replacement of lamp
(only for systems using a spectrophotometer with an halogen lamp)
WARNING:
RISK OF BURNS
Do not handle the lamp to be replaced immediately after turning off the analyser.
Before handling, allow the lamp to cool for approximately 10 minutes after switching
the analyser off.
•
Open the lid and insert the lid support tool (see pages 35 and 36).
•
Switch the analyser off and unplug the power cable.
•
Remove the bottom partition at the rear of the rinsing well.
•
Disconnect the lamp from its supply.
•
Unscrew the securing screws located on both sides of the lamp in order to remove the lamp-support unit.
•
Remove the old lamp.
WARNING:
RISK OF BURNS
Do not touch the lamp with your fingers.
•
Wipe the new lamp with a soft cloth.
!
Do not touch the lamp with your fingers.
•
Put the new lamp and its base in place, with the guide pin positioned downwards.
•
Tighten the securing screws using a screwdriver.
•
Connect the lamp to its power supply.
•
Replace the partition.
•
Remove the lid support tool and close the lid (see page 36).
•
Switch the analyser back on and start it up.
Maintenance
5-4- Analyser Interventions
IDS-iSYS User Manual - Revision M1!
Maintenance 5-4
126
Software version V 14