R|1|^^^25OHD|25,2|ng/mL|10,0 to 100,0|N||F||||20111124133726|IDS-iSYS B300B0378|
This block represents the information sent through by the IDS-iSYS to the LIS after a test has been carried out
for a patient.
Here are the various fields of this block used in the protocol version:
* The field value corresponds to the above block
ASTM Name of the field
Value*
Description and use in IDS-iSYS system
Record Type ID
R
Fixed value R for indicating a Result block.
Sequence Number
1
The number of the result within the Patient block
begins at 1 and is increased for each Result block
in the message.
Universal Test ID
^^^25OHD
This ASTM field consists of 4 components:
Universal test ID.
Universal test ID name.
Universal test ID type.
Manufacturer’s code.
Only the last component is used and will be
provided with information with the unique IDS-
iSYS identifier of the analyte.
Data or Measurement Value
25,2
The value in the form of ASCII chain.
Results which cannot be calculated are sent in the
format * * * * * . Each character * is separated by a
space.If the value is outside the measuring range,
is transmitted in this field:
* * * * * when the result is below the measuring
range
* * * * * when the result is above the measuring
range
(For 25OHD: <5 or >140)
Units
ng/mL
Blocks used
Reference Range
10,0 to 100,0
Normal reference values used for the patient sent
through in the form of an ASCII chain.
Result abnormal Flags
N
This field may take the values:
L = Below low normal.
H = Above high normal.
LL = Below alarm low.
HH = Above alarm high.
< of absolute low (Field of measurement).
> of absolute high (Field of measurement).
N = Normal.
A = Abnormal.
U = Significant change Up.
D = Significant change Down.
B = Better.
W = Worse.
Only A and N are managed by IDS-iSYS.
2- ASTM Compatible Version
2-5- The Result Block
IDS-iSYS Connection Protocol - Revision N1"
2- ASTM Compatible Version
" 12
Software version V 14

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the IDS IDS-iSYS and is the answer not in the manual?

IDS IDS-iSYS Specifications

General IconGeneral
BrandIDS
ModelIDS-iSYS
CategoryMeasuring Instruments
LanguageEnglish