R|1|^^^25OHD|25,2|ng/mL|10,0 to 100,0|N||F||||20111124133726|IDS-iSYS B300B0378|
This block represents the information sent through by the IDS-iSYS to the LIS after a test has been carried out
for a patient.
Here are the various fields of this block used in the protocol version:
* The field value corresponds to the above block
Description and use in IDS-iSYS system
Fixed value R for indicating a Result block.
The number of the result within the Patient block
begins at 1 and is increased for each Result block
in the message.
This ASTM field consists of 4 components:
•
Universal test ID.
•
Universal test ID name.
•
Universal test ID type.
•
Manufacturer’s code.
Only the last component is used and will be
provided with information with the unique IDS-
iSYS identifier of the analyte.
Data or Measurement Value
The value in the form of ASCII chain.
Results which cannot be calculated are sent in the
format * * * * * . Each character * is separated by a
space.If the value is outside the measuring range,
is transmitted in this field:
•
* * * * * when the result is below the measuring
range
•
* * * * * when the result is above the measuring
range
(For 25OHD: <5 or >140)
Normal reference values used for the patient sent
through in the form of an ASCII chain.
This field may take the values:
•
L = Below low normal.
•
H = Above high normal.
•
LL = Below alarm low.
•
HH = Above alarm high.
•
< of absolute low (Field of measurement).
•
> of absolute high (Field of measurement).
•
N = Normal.
•
A = Abnormal.
•
U = Significant change Up.
•
D = Significant change Down.
•
B = Better.
•
W = Worse.
Only A and N are managed by IDS-iSYS.
2- ASTM Compatible Version
2-5- The Result Block
IDS-iSYS Connection Protocol - Revision N1"
2- ASTM Compatible Version
" 12
Software version V 14