3-3-3- Installation of IDS-iSYS cuvettes
WARNING:
RISK OF INJURY
Do not place your hand or fingers into the loader.
If necessary, remove the empty IDS-iSYS Cuvettes box.
•
Scan the replacement box with the barcode reader.
•
Within 10 seconds of the barcode reading, install the IDS-iSYS Cuvettes cube into the loader with the
window to the front.
•
Remove the protective plastic strip: pull forwards to break the seal. Gently pull upwards to remove the
protective strip. Place the protective strip over the cube to protect the window from dust.
•
If the barcode is not read by the barcode reader, click on the ancillary reagent then use the keyboard to enter
the barcode identifier.
Use
3-3- Installation of Ancillary Reagents (continued)
IDS-iSYS User Manual - Revision M1!
Use 3-3
50
Software version V 14