5-4-13- Repeat an automatic maintenance
Certain maintenance activities are carried out automatically by the system, without user’s intervention.
If a fault prevents the automatic maintenance being carried out in full, an error message is displayed. A
Warning message remains in the WARNING window until a new maintenance is complete without any error.
When this maintenance cannot be carried out automatically by the system, a new maintenance can be
requested by the user.
5-4-13-1- Washer needle cleaning
This automatic maintenance takes place when the system is placed in standby mode.
After cleaning the sampling probe using D-SORB solution, four cuvettes are prepared, each filled with 500 µl of
this solution (injected in two steps). The cuvettes are then transferred to the four washers: the aspiration needle
of the washer is cleaned when aspirating the D-SORB solution. This step is followed by a washing step. This
maintenance takes around 3 minutes et uses 3 ml de solution D-SORB.
Check that D-SORB and WASH S solutions are present on the system, both with sucient volumes..
From the MAINTENANCE menu, select REPEAT AN AUTOMATIC MAINTENANCE, then select:
As soon as the menu option is selected, the maintenance is carried out by the system.
When maintenance is completed without any error, the date and time are recorded in the journal of events (in
the SYSTEM RECORDS tab).
5-4-13-2- Cleaning the probe
This automatic maintenance takes place when the system is placed in standby mode.
Certain assays performed on the system require a daily cleaning of the sampling probe with the
Immunocleaner solution. This product is supplied in the form of a reagent cartridge and must be installed in
the reagent compartment.
During this maintenance, the probe performs three times the following cycle : pipetting 300 µl of
Immunocleaner solution, waiting for 30 seconds above the rinsing well, rinsing. The final rinsing uses D-Sorb
solution.
Check that IDS-iSYS D-Sorb and Immunocleaner solutions are present in sucient quantities on the
analyser.
From the MAINTENANCE menu, select REPEAT AN AUTOMATIC MAINTENANCE, then select:
As soon as the menu option is selected, the maintenance is carried out by the system.
Maintenance
5-4- Analyser Interventions (continued)
IDS-iSYS User Manual - Revision M1!
Maintenance 5-4
145
Software version V 14