Configurable options (continued):
Active for the option «Clot Detection»
When a clot is detected in a sample, results from this
sample’s profile that have already been transferred to the
LIS will be resent (see 7- Important notes, page 67).
Active for the option «Clot Detection»
Results that have already been transferred to the LIS are
not resent when a clot is detected in a sample.
The field «Specific ancillaries» in the Manufacturer block will
not contain the components corresponding to the
traceability of AP SUBSTRATE and washer(s) used in the
pretreatment step.
The relevant field is described in page 31.
The field «Specific ancillaries» in the Manufacturer block will
contain the components corresponding to the traceability of
AP SUBSTRATE and washer(s) used in the pretreatment
step.
The relevant field is described in page 31.
3-10-3- Transferring quality control results
To tr an sf er th e qu al it y c on tro l re su lt s, s el ec t th e c or res po nd in g o pt io n, i n t he t ab
under SET-UP/SETTINGS/SYSTEM menu:
3- ASTM Compatible V2 Version
3-10- Line of Communication (continued)
IDS-iSYS Connection Protocol - Revision N1"
3- ASTM Compatible V2 Version
" 37
Software version V 14