Close the drawer.
•
Open the lid and insert the lid support tool (see pages 35 and 36).
•
Clean the Plexiglas
®
over the tray with absorbent paper soaked in de-ionised water. Then use a disinfectant
wipe/solution suitable for medical devices.
•
Remove the lid support tool and close the lid (see page 36).
•
When maintenance is complete, click on the corresponding , then confirm the
maintenance by clicking on YES.
5-2-1-3- Cleaning the rinsing well
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
This part of the analyser is in contact with biological samples. There is, therefore, a
potential risk of infection.
Wear disposable gloves for all handling procedures.
•
Open the lid and insert the lid support tool (see pages 35 and 36).
•
Pour alcohol at 70% v/v (ethyl or isopropyl) into the rinsing well.
•
Use a cotton bud soaked in alcohol at 70% v/v (ethyl or isopropyl) to clean the well.
•
When maintenance is complete, click on the corresponding , then confirm the
maintenance by clicking on YES.
5-2-1-4- Checking dilutors and IDS-iSYS System Liquid pumps
•
Remove the cover from the right-hand hand side of the analyser.
•
Check for any leaks on the tubing connections at the outlet of each dilutor.
•
Check for any leaks on the IDS-iSYS System Liquid pumps.
•
Replace the cover.
•
When maintenance is complete, click on the corresponding , then confirm the
maintenance by clicking on YES.
This maintenance activity can be carried out with the lid opened.
Maintenance
5-2- Weekly Maintenance (continued)
IDS-iSYS User Manual - Revision M1!
Maintenance 5-2
120
Software version V 14