5-1-1-3- Cleaning needle exterior
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
This part of the analyser is in contact with biological samples. There is, therefore, a
potential risk of infection.
Wear disposable gloves for all handling procedures.
Open the lid and insert the lid support tool (see pages 35 and 36).
Gently clean the probe with a cloth soaked in deionized water to remove traced on the needle.
Then use absorbent paper soaked in alcohol at 70% v/v (ethyl or isopropyl) to decontaminate the needle.
When the cleaning procedure is finished, verify the tightness of the needle.
Remove the lid support tool and close the lid (see page 36).
!
Handle the probe with care. Do not twist or bend the probe during cleaning.
5-1-1-4- Decontamination of the probe
!
This maintenance activity is managed only for systems using DSORB solution.
For other systems, this maintenance activity is not required.
Check that IDS-iSYS D-Sorb solution is present on the analyser.
Click on the corresponding button in front of this item in the maintenance list.
Click . The probe will automatically aspirate the IDS-iSYS D-Sorb solution.
When maintenance is completed without any error, the date and time it is carried out are recorded.
Maintenance
5-1- Daily Maintenance (continued)
IDS-iSYS User Manual - Revision M1!
Maintenance 5-1
117
Software version V 14