5-4-5- Replacement of IDS-iSYS Wash Solution pump (Immunoassay)
Switch the analyser o.
Remove the cover from the right-hand side of the analyser.
Press on the two pins at each side of the pump to be changed and remove the pump body from its shaft.
Disconnect the pipes from the connectors and plug into the new pump.
Clean the shaft of the pump with absorbent paper soaked in alcohol at 70% v/v (ethyl or isopropyl).
Put the pump back in place on its shaft.
Replace the cover of the analyser.
Switch the analyser back on again and perform the start-up procedure.
Carry out partial priming of the washers in order to prime the tubing circuit again.
5-4-6- Replacement of IDS-iSYS System Liquid pump
Switch the analyser o.
Remove the cover from the right-hand side of the analyser.
Press on the two pins at each side of the pump to be changed and remove the pump body from its shaft.
Disconnect the pipes from the connectors and plug into the new pump.
Clean the shaft of the pump with absorbent paper soaked in alcohol at 70% v/v (ethyl or isopropyl).
Put the pump back in place on its shaft.
Replace the cover of the analyser.
Switch the analyser back on again and perform the start-up procedure.
Carry out partial priming of the arm in order to prime the circuit again.
5-4-7- Replacement of liquid waste pump
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
This part of the analyser is in contact with biological samples. There is, therefore, a
potential risk of infection.
Wear disposable gloves for all handling procedures.
Switch the analyser o.
Remove the cover from the right-hand side of the analyser.
Press on the two pins at each side of the pump to be changed and remove the pump body from its shaft.
Disconnect the pipes from the connectors and plug into the new pump.
Clean the shaft of the pump with absorbent paper soaked in alcohol at 70% v/v (ethyl or isopropyl).
Put the pump back in place on its shaft.
Replace the cover of the analyser.
Switch the analyser back on again and perform the start-up procedure.