5-5-1- Decontaminating the containers of solid and liquid waste
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
This part of the analyser is in contact with biological samples. There is, therefore, a
potential risk of infection.
Wear disposable gloves for all handling procedures.
Solid waste tray
Empty and decontaminate the waste tray once emptied with a disinfectant solution suitable for medical
devices.
Rinse it with water, then decontaminate again with a disinfectant solution suitable for medical devices.
Dry the tray prior to replacing in its compartment.
Liquid waste container
Rinse the container with freshly prepared bleach once the liquid waste has been disposed of.
Then rinse with tap water.
Add the appropriate volume of bleach prior to reusing the container with the analyser (see Section
Appendices, A-1- Waste Disposal, page A2).
Maintenance
5-5- Analyser Cleaning (continued)
IDS-iSYS User Manual - Revision M1!
Maintenance 5-5