Calibration status
The icon on the cartridge shows the status of the current calibration.
Not calibrated
Valid Calibration
Calibration accepted by the user
Expired or not valid Calibration
Biochemistry reagents
The following information is displayed for Biochemistry reagents:
test name,
lot number.
3-2-3- Management of on board reagents
Reagents are managed by the system in terms of:
available on board quantity (number of tests for Immunoassay cartridges and volume for BIochemistry
reagents),
expiry date,
in-use stability,
on board stability.
The expiry date is managed for the reagent lot. Other data are individually managed per cartridge or per bottle
of the lot.
When loading a new cartridge or a new reagent bottle the system records the date and time of installation.
All operations of loading / unloading are recorded for each cartridge or each reagent bottle.
Cumulative on board time is calculated from these data (displayed in the format: Days \ Hours : Minutes :
Seconds)
All this information can be displayed by clicking on the position occupied by the reagent and is stored in the
menu MANAGEMENT OF LOTS / REAGENTS.
The two following stabilities, included in the lot data, are applied to each cartridge or reagent bottle:
in-use stability: period of time during which a cartridge or a reagent bottle can be used after opening and
storage in conditions defined in reagent IFU. In-use stability is defined by the system from the date and time
of first installation of the cartridge or reagent bottle into the reagent compartment.
on board stability: period of time during which a cartridge or a reagent bottle can be used after opening
when stored on-board. On-board stability is defined by the system from the date and time of first installation
of the cartridge or reagent bottle and takes into account the cumulative on board time of the cartridge or
reagent bottle.
According to the nature of the reagents, these stabilities are expressed in days or in hours.
Use
3-2- Installation of Reagents (continued)
IDS-iSYS User Manual - Revision M1!
Use 3-2