3-6-2- Fixed tray
!
Take care when opening and closing the sample drawer: rough handling when opening and
closing the drawer may cause the internal samples to spill.
If the light is green, open the sample compartment.
Installation using a barcode
Scan the barcode with the reader located on the front face. The analyser will beep when the barcode has
been correctly read.
Within 10 seconds, place in a free position on the rack.
If the barcode is recognised, the analyser will beep a second time.
On the interface, the display will indicate that the position has been identified.
If no second beep is heard, the position is considered to be occupied but not identified. In this case, re-start
the barcode reading and installation process.
If the barcode label is illegible,remove the tube and click on .
Use the keyboard to enter the barcode identifier.
Select ‘Vial association with samples tray’ and click ‘OK’.
Click on .
Within 10 seconds, place in a free position on the rack.
If the association process is correct, the analyser will beep .
On the interface, the display will indicate that the position has been identified.
For tubes from patients whose profile has already been sent through via a centralised computer or manually
programmed, the profile in the memory is automatically associated with the position.
If the profile is not in the memory, the profile programming window opens (see Section 3-7, page 69).
!
For the calibrators and controls supplied by IDS, enter the 12 figures from the barcode: the
product, the lot number, the expiry date as well as the table of corresponding values are
automatically entered.
Installation without barcode
Place the tube(s) or sample cup(s) in a free position on the tray.
Click on the position which is occupied but not identified.
The software opens a window which allows the position to be configured.
Then configure the position (see 3-6-3- Configure a position without barcode, page 65).
Use
3-6- Loading of Calibrators, Controls or Samples in the Sample Compartment (continued)
IDS-iSYS User Manual - Revision M1!
Use 3-6
64
Software version V 14