P|1||20105863||BONTEMPS^Robert||19500226|M|||||DUPONT|||||Comment||||||||||||||
Radiology||TANNER STAGE^5
This block represents the information relating to a patient.
Here are the various fields of this block used in the protocol version:
* The field value corresponds to the above block
Description and use in IDS-iSYS system
Fixed value P to indicate a Patient block.
Number of the Patient block in the message starts
at 1 and is increased for each block in the
message.
Practice Assigned Patient ID
Laboratory Assigned Patient
ID
Field consisting of 5 components last name, first
name, initials, suffix and title.
Only the first 2 fields are used.
When sending a control from IDS-iSYS to LIS, the
first component in the field will be used for the
name of control.
Date of birth in format YYYYMMDD.
Reference of patient values:
•
M for Man.
•
F for Woman.
•
E for Child.
•
U for others.
In the direction LIS=>IDS-iSYS:
•
M for Man.
•
F for Woman.
•
U for Child.
•
U for others.
4- ASTM Compatible V3 Version
4-3- The Patient Block
IDS-iSYS Connection Protocol - Revision N1"
4- ASTM Compatible V3 Version
" 41
Software version V 14