1-4- Installation
The packaging of the has been designed to prevent any damage occurring during
transportation.
The analyser can be stored in its original packaging under the following conditions:
Storage temperature " " 10-40°C.
Relative humidity ! ! 80% non condensing.
In the event of prolonged storage, the analyser performance must be checked. The unpacking, installation and
initial qualification of the analyser must only be carried out by a qualified IDS representative.
!
After installation by Technical Services or an IDS representative, do not handle or move the
analyser.
If the analyser must be moved, contact Technical Services or IDS representative.
1-4-1- Environment
A satisfactory installation site is essential for the analyser to function correctly.
The user must ensure compliance with the conditions required in terms of environment and electricity supply in
order to maintain the performance of the analyser and to guarantee safe use for the operator.
Environmental conditions required:
The analyser must not be exposed to direct light.
A clean and ventilated air environment.
The analyser must be placed on a flat work surface, capable of supporting its weight (103 kg) without
significant vibration.
The surrounding temperature must be between 15°C and 30°C.
Relative humidity must be within 20 and 80% (non-condensing).
The analyser must not be installed under an air-conditioning unit.
Clearance of at least 15 cm must be provided at the rear, front, left and right of the analyser to allow
evacuation of heat produced by the apparatus.
The analyser must be installed in such as way as to allow the user to easily access the ON/OFF button and
the main cable inlet.
The analyser must not be installed near strong sources of electromagnetic radiation and electrical
interference (e.g. refrigerators).
The IVD medical device complies with the emission and immunity requirements described in the standard
IEC61326-2-6.
Physical characteristics of the analyser:
Dimensions of the analyser! L 105 x H 70 x W 75 cm.
Overall dimensions ! ! L 124 x H 110 x W 71 cm.
Weight ! ! ! ! 103 kg.
Basis weight! ! ! 130 kg/m
2
*
Computer system ! ! L 60 x H 40 x W 50 cm.
* Analyser only. Basis weight of analyser and its special table: 278 kg/m
2
Operating principle
1-4- Installation
IDS-iSYS User Manual - Revision M1"
Operating principle 1-4
10
Software version V 14