L|1|N
This block is used in order to indicate the end of an ASTM block. It consists of the following fields:
* The field value corresponds to the above block
ASTM Name of the field
Value*
Description and use in IDS-iSYS system
Record Type ID
L
Fixed value L for indicating a Terminator block.
Sequence Number
1
Always at 1 as there is only one sequence per
message.
Termination code
N
This code designates of end of the message with
the following values, depending on the situation:
N = Normal.
T = Abandon issuer.
R = Abandon receiver.
E = Unknown Error.
Q = Error in last request.
I = No Information available for last query.
F = Last request for information processed. of
comment.
4- ASTM Compatible V3 Version
4-9- The Message Terminator Block
IDS-iSYS Connection Protocol - Revision N1"
4- ASTM Compatible V3 Version
" 53