5-3-2-2- Rinsing of the AP Substrate tubing
Prepare a 500 mL bottle containing de-ionised water to mimic the AP Substrate.
Click on the corresponding button in front of this item in the maintenance list:
Replace the AP Substrate bottle with the de-ionised water bottle.
Click on .
When the rinsing is complete, the following message displays:
Remove the de-ionised water bottle and replace the original AP Substrate bottle.
Click on .
When maintenance is completed without any error, the date and time are recorded.
Maintenance
5-3- Monthly Maintenance (continued)
IDS-iSYS User Manual - Revision M1!
Maintenance 5-3