1-3- Characteristics
Analyser
•
Multidiscipline Immunoassay and Biochemistry analyser.
•
Continuous loading.
•
Configuration:
•
Immunoassay (4 washers) + Biochemistry.
Analyser Physical
Characteristics
•
Analyser dimensions:! L 105 cm x H 70 cm x W 75 cm.
•
Overall dimensions: ! L 124 cm x H 110 cm x W 71 cm.
•
Weight: ! ! ! 103 kg.
•
Basis weight: 130 kg/m
2
•
Sound level:! ! 55 dB (average in run cycle); up to 66 dB
•
Emitted heat: up to 530 W (1808 BTU/h or 16,5 cal/s)
•
Computer System: ! L 60 cm x H 40 cm x W 50 cm.
•
Immunoassay! ! Up to 120 tests/hour (Assay dependent)*.
•
Biochemistry! ! 240 tests/hour.
* According to configuration.
•
15 Immunoassay.
•
Up to 50 Biochemistry.
•
Chemiluminescence (luminometer).
•
Enzymatic assays (spectrophotometer)
•
Colorimetric and enzymatic assays.
•
Turbidimetric assays.
•
Serum, plasma or urine.
•
Primary tubes 5 mL (13 x 75 mm), 7 mL (13 x 100 mm).
10 mL (16 x 75 mm).
•
Secondary tubes (13 x 75 mm).
•
Cups (ref. IS-CSC105 and ref. IS-CSC115).
•
Depending on the configuration, positions for samples, calibrators and
controls:
•
Fixed tray : 64 positions
•
Removable trays : 20 positions per tray. Up to 6 trays on board.
Possibility of special tray for calibrator and control vials (16 positions
for 2ml vials, diameter 18 mm).
•
Liquid level detection by capacitance.
•
Clot detection.
•
Dilutions and automatic pre-treatments.
•
Reader on front face for barcode identification of cuvettes and ancillary
reagents. Used for sample identification in fixed tray configuration.
•
Integrated reader for reagents barcode
•
For configuration with removable trays, reader integrated into the
sample compartment.
Operating principle
1-3- Characteristics
IDS-iSYS User Manual - Revision M1"
Operating principle 1-3
7
Software version V 14