Q|1|^7148344||ALL||||||||O
This block represents a request for information whose origin now can only be the IDS-iSYS.
This request will be used by the IDS-iSYS after discovering a tube for a patient whose profile is
known on the work list in order to request the tests to be carried out on this tube.
Here are the various fields of this block used in the protocol version:
* The field value corresponds to the above block
Description and use in IDS-iSYS system
Fixed value Q to indicate a Request Information block.
The request number starts at 1 and is increased for
each request block in the message.
This field consists of 3 elements; only the first 2 are
used by the IDS-iSYS in order to specify the PID
(optional) and the SID (obligatory) allowing the profile
quoted to be identified.
This field can either indicate 1 particular test (in the
identifier via its ID cf Test Order block) or request all
the tests via the code ALL.
Nature of request Time
Limit
Not used in the direction IDS-iSYS=>LIS.
Useful if LIS => IDS-iSYS with the value R to say that
the date to be taken into account in the event of filter
is the date of the test results and the date of arrival of
the profile (code S).
Beginning request Results
Date and Time
Not used
Useful in the direction LIS=>IDS-iSYS.
Ending request Results
Date and Time
Not used
Useful in the direction LIS=>IDS-iSYS.
Requesting Physician
Name
Requesting Physician
Telephone Number
3- ASTM Compatible V2 Version
3-8- The Request Information Block
IDS-iSYS Connection Protocol - Revision N1"
3- ASTM Compatible V2 Version
" 32
Software version V 14