Remove the activation key. The removal is detected by the analyser and the message is displayed:
•
Remove the lid support tool and close the lid (see page 36). The lid closing is detected by the analyser and
the main position adjustment window is displayed.
5-4-11- Unclogging the sampling probe
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
The probe is in contact with biological samples. There is, therefore, a potential risk of
infection.
Wear disposable gloves for all handling procedures.
The probe must be disposed of in accordance with current local regulations.
!
Use only the nylon mandrel provided for this purpose with the analyser.
Do not use a metallic mandrel that will damage the internal surface of the probe.
•
Open the lid and insert the lid support tool (see pages 35 and 36).
•
If the lid of the analyser is equipped with a locking system, the sampling arm moves to its home position.
•
Install the specific protective tray under the probe.
•
Unscrew the sampling probe.
•
Insert the mandrel into the bottom of the probe, and pass it through the entire length of the probe, until the
probe becomes unblocked.
•
Outside of the analyser, verify that the probe is fully unblocked by passing deionized water through the top of
the probe with a syringe:
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
The probe is in contact with biological samples. There is, therefore, a potential risk of
infection.
Wear disposable gloves for all handling procedures.
The probe must be disposed of in accordance with current local regulations.
Maintenance
5-4- Analyser Interventions (continued)
IDS-iSYS User Manual - Revision M1!
Maintenance 5-4
141