If the probe is properly unblocked, the water will exit the probe in a straight jet and the syringe plunger can
be pushed freely.
•
If the syringe plunger must be strongly pushed and if water drips when exiting the probe, repeat this
procedure.
•
Then clean the probe with an absorbent paper soaked in alcohol at 70% v/v (ethyl or isopropyl).
•
Reassemble the probe by screwing it in by hand as far as possible to avoid risk of leakage.
•
Remove the protective tray.
•
Remove the lid support tool and close the lid (see page 36).
When the analyser is not in cycle mode:
•
Adjust the probe reference position (see page 132 and following).
•
Prime the fluidics circuit as follows:
•
Select ARM-Mixed Partial Priming, then click on .
The analyser starts up the fluidics circuit and rinses the sampling needle.
Maintenance
5-4- Analyser Interventions (continued)
IDS-iSYS User Manual - Revision M1!
Maintenance 5-4
142
Software version V 14