If the probe is properly unblocked, the water will exit the probe in a straight jet and the syringe plunger can
be pushed freely.
If the syringe plunger must be strongly pushed and if water drips when exiting the probe, repeat this
procedure.
Then clean the probe with an absorbent paper soaked in alcohol at 70% v/v (ethyl or isopropyl).
Reassemble the probe by screwing it in by hand as far as possible to avoid risk of leakage.
Remove the protective tray.
Remove the lid support tool and close the lid (see page 36).
When the analyser is not in cycle mode:
Adjust the probe reference position (see page 132 and following).
Prime the fluidics circuit as follows:
Select ARM-Mixed Partial Priming, then click on .
The analyser starts up the fluidics circuit and rinses the sampling needle.
Maintenance
5-4- Analyser Interventions (continued)
IDS-iSYS User Manual - Revision M1!
Maintenance 5-4
142
Software version V 14