R|1|^^^25OHD^04^F^1|63,9|ng/mL|10,0 to 100,0|N||F|||SYSTEM^SYSTEM|
20111123171733|IDS-iSYS B300B0378
This block represents the information sent through by the IDS-iSYS to the LIS after a test has been carried out
for a patient.
Here are the various fields of this block used in the protocol version:
* The field value corresponds to the above block
Description and use in IDS-iSYS system
Fixed value R for indicating a Result block.
The number of the result within the Patient block
begins at 1 and is increased for each Result block
in the message.
This ASTM field consists of 8 components:
- Universal test Id : forces NULL
- Universal test Id Name: forces NULL.
- Universal test Id Type: forces NULL.
-Manufacturer’s Code: contains the unique
IDS-iSYS identifier for the analyte.
-Test type: 02 for a control, 04 for a patient
-Result type: F for final, P for partial.
-Replicate number: 1, 2,up to 10
-Control identifier, defined as follow:
•
21 for a control Level 1
•
22 for a control Level 2
•
23 for a control Level 3
•
24 for a control Level 4
For a patient,this component is NULL.
-Dilution rate: NULL if not applicable.
Data or Measurement Value
The value in the form of ASCII chain.
Results which cannot be calculated are sent in the
format * * * * * . Each character * is separated by a
space.If the value is outside the measuring range,
is transmitted in this field:
•
< Low limit of the sample measurement range,
when the result is below the measuring range
•
> High limit of the sample measurement range,
when the result is above the measuring range
(For 25OHD: <5 or >140)
Normal reference values used for the patient sent
through in the form of an ASCII chain.
4- ASTM Compatible V3 Version
4-5- The Result Block
IDS-iSYS Connection Protocol - Revision N1"
4- ASTM Compatible V3 Version
" 46
Software version V 14