ASTM Name of the field
Value*
Description and use in IDS-iSYS system
Result abnormal Flags
N
This field may take the values:
L = Below low normal.
H = Above high normal.
LL = Below alarm low.
HH = Above alarm high.
< of absolute low (Field of measurement).
> of absolute high (Field of measurement).
N = Normal.
A = Abnormal.
U = Significant change Up.
D = Significant change Down.
B = Better.
W = Worse.
L, H, <, >, A and N are managed by IDS-iSYS.
All associated messages are sent in a Comment
block.
Nature of Abnormality Testing
S
This field forces S or A for the IDS-iSYS can take
the values:
A = Age.
S = Sex.
R = Race.
And is able to provide an indication about the
bases on which the values of normality were
determined.
Result Status
F
This field forces to F or P (replicate) on the IDS-
iSYS can take the following values:
C = Correction.
P = Preliminary results.
F = Final results.
X = Request not honored.
I = Results pending.
S = Partial results.
Date of change in Instrument
Normative Values
Not used.
Operator ID
SYSTEM^SYSTEM
Identifying the operators who handled the profile.
This field consists of 2 components:
- Identifier of the operator who creates the
profile; if the profile was transferred by the LIS,
the identifier SYSTEM is sent.
- Identifier of the operator who stores/tranfers
the profile; if the profile was automatically
stored/transferred, the identifier SYSTEM is
sent.
Date and Time Test Started
Not used.
Date and Time Test
Completed
20111123171733
Date and time of result of test on the GUI.
Instrument Identification
IDS-iSYS B300B0378
Name of device + Serial number.
4- ASTM Compatible V3 Version
4-5- The Result Block (continued)
IDS-iSYS Connection Protocol - Revision N1"
4- ASTM Compatible V3 Version
" 47
Software version V 14