Description and use in IDS-iSYS system
This field may take the values:
•
L = Below low normal.
•
H = Above high normal.
•
LL = Below alarm low.
•
HH = Above alarm high.
•
< of absolute low (Field of measurement).
•
> of absolute high (Field of measurement).
•
N = Normal.
•
A = Abnormal.
•
U = Significant change Up.
•
D = Significant change Down.
•
B = Better.
•
W = Worse.
L, H, <, >, A and N are managed by IDS-iSYS.
All associated messages are sent in a Comment
block.
Nature of Abnormality Testing
This field forces S or A for the IDS-iSYS can take
the values:
•
A = Age.
•
S = Sex.
•
R = Race.
And is able to provide an indication about the
bases on which the values of normality were
determined.
This field forces to F or P (replicate) on the IDS-
iSYS can take the following values:
•
C = Correction.
•
P = Preliminary results.
•
F = Final results.
•
X = Request not honored.
•
I = Results pending.
•
S = Partial results.
Date of change in Instrument
Normative Values
Identifying the operators who handled the profile.
This field consists of 2 components:
- Identifier of the operator who creates the
profile; if the profile was transferred by the LIS,
the identifier SYSTEM is sent.
- Identifier of the operator who stores/tranfers
the profile; if the profile was automatically
stored/transferred, the identifier SYSTEM is
sent.
Date and Time Test Started
Date and Time Test
Completed
Date and time of result of test on the GUI.
Name of device + Serial number.
4- ASTM Compatible V3 Version
4-5- The Result Block (continued)
IDS-iSYS Connection Protocol - Revision N1"
4- ASTM Compatible V3 Version
" 47
Software version V 14