A-2- Decontaminating the Analyser
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
These parts of the analyser are in contact with biological samples.
There is, therefore, a potential risk of infection.
Wear disposable gloves for all handling procedures.
The analyser must be decontaminated after carrying out the regular maintenance procedures described in
Section 5 of this user manual. It is essential that decontamination is carried out:
Prior to any intervention by the Technical Services & Support personnel.
Prior to any transportation of the system.
A cleaning/decontaminating declaration must be completed by the user.
The declaration is printed on the following page and should be duplicated and, once completed, attached to
the analyser in a prominent position.
It is essential for this declaration to accompany the analyser during any transportation (e.g. return to factory).
Cleaning/decontamination process:
Decontaminate the probe with a wipe soaked in a decontaminating solution suitable for medical devices.
Decontaminate the sample tray and reagent compartment with a wipe soaked in a decontaminating solution
suitable for medical devices.
Decontaminate the rinsing well and the liquid waste tubing by pouring bleach (commercial preparation) into
the drainage hole in the rinsing well.
Empty the solid waste tray and liquid waste container.
Decontaminate the solid waste tray with a wipe soaked in a decontaminating solution suitable for medical
devices.
Decontaminate the bodywork, the keyboard and the keys of the computer with a wipe soaked in a
decontaminating solution suitable for medical devices.
!
Do not use any spray products on the analyser.