5-2-2- Immunoassay Maintenance
5-2-2-1- Rinsing of the IDS-iSYS Triggers tubing
•
Open the lid and insert the lid support tool (see pages 35 and 36).
•
Prepare two 250 mL bottles containing de-ionised water to mimic the IDS-iSYS Triggers.
•
Click on the corresponding button in front of this item in the maintenance list.
•
Follow the on-screen instructions.
•
When maintenance is completed without any error, the date and time are recorded.
This procedure can be carried out with the lid opened.
When this maintenance is completed, proceed to the weekly shutdown of the computer:
•
From the File menu, click on EXIT, the software will close.
•
Then turn off the computer following the procedure for shutting down WINDOWS.
•
Switch the computer on, then open the software by double-clicking on the IDS-iSYS icon.
•
Once the software is open, enter your access code to open a session.
Maintenance
5-2- Weekly Maintenance (continued)
IDS-iSYS User Manual - Revision M1!
Maintenance 5-2
121
Software version V 14