5-2-2- Immunoassay Maintenance
5-2-2-1- Rinsing of the IDS-iSYS Triggers tubing
Open the lid and insert the lid support tool (see pages 35 and 36).
Prepare two 250 mL bottles containing de-ionised water to mimic the IDS-iSYS Triggers.
Click on the corresponding button in front of this item in the maintenance list.
Follow the on-screen instructions.
When maintenance is completed without any error, the date and time are recorded.
!
This procedure can be carried out with the lid opened.
When this maintenance is completed, proceed to the weekly shutdown of the computer:
From the File menu, click on EXIT, the software will close.
Then turn o the computer following the procedure for shutting down WINDOWS.
Switch the computer on, then open the software by double-clicking on the IDS-iSYS icon.
Once the software is open, enter your access code to open a session.
Maintenance
5-2- Weekly Maintenance (continued)
IDS-iSYS User Manual - Revision M1!
Maintenance 5-2
121
Software version V 14