ASTM Name of the field
Value*
Description and use in IDS-iSYS system
Request Information
Status Code
O
Potential values:
C
P
F
X
I
S
M
R
A
N
O
D
Only the values O and X are managed in the
direction IDS-iSYS to LIS.
A is used when the LIS does not respond to a
request. In such cases, the request is cancelled
(this field takes the value A). The request is
transmitted 3 times, after which the transmission is
aborted.
3- ASTM Compatible V2 Version
3-8- The Request Information Block (continued)
IDS-iSYS Connection Protocol - Revision N1"
3- ASTM Compatible V2 Version
" 33
Software version V 14

Table of Contents