Installation of AP Substrate
Remove the cap from the new ancillary reagent bottle.
If necessary, take out the bung and remove the empty bottle.
Scan the ancillary with the barcode reader.
The analyser will beep for the first time.
Within 10 seconds, install the new bottle in its position.
The analyser will beep a second time.
Install the supply line into the bottle.
If the barcode is not read by the barcode reader, click on the ancillary reagent then use the keyboard to enter
the identifier.
Once a new bottle is installed, a message is displayed:
Click on to confirm the priming of the AP substrate.
The analyser starts the AP Substrate priming sequence.
Comment: The automatic priming of the AP Substrate circuit can be cancelled by clicking on . In
this case, the priming will take place automatically when starting a run cycle.
Use
3-3- Installation of Ancillary Reagents (continued)
IDS-iSYS User Manual - Revision M1!
Use 3-3