Description and use in IDS-iSYS system
This field consists of two sections:
•
The sample type which can take for the IDS-
iSYS the following values:
•
U = Urine.
•
S = Serum.
•
O = Other.
•
The sample source (place where the sample
was taken) is not used.
Physician Telephone
Number
Date and Time Result
Reported or Last Modified
Instrument Charge to
Computer System
This field may take the values:
•
O = Order.
•
C = Correction.
•
P = Preliminary results.
•
F = Final results.
•
X = Cancel.
•
I = Instrument pending (in response to a
request).
•
Y = No order on record for this test
(in response to a request).
•
Z = No record for this patient
(in response to a request).
•
Q = Response to a request.
In the device, C is managed when a clot is
detected on a sample (see page 67).
Location or Ward of
Specimen Collection
Nosocomial Infection Flag
2- ASTM Compatible Version
2-4- The Order Block (continued)
IDS-iSYS Connection Protocol - Revision N1"
2- ASTM Compatible Version
" 11