Description and use in IDS-iSYS system
This field may contain the values:
•
S = STAT.
•
A = ASAP.
•
R = Routine.
•
C = Callback.
•
P = Preoperative.
For the device, only S has any influence, the
passage of the associated profile into Urgent, and
the other values are ignored.
Requested/Ordered
Date and Time
Specimen Collection
Date and Time
Action to be carried out on the block test or tests.
This field may contain the values:
•
A for Add.
•
C for Cancel.
•
N for New.
•
P for Pending.
•
Q for QC.
For the device, A and N are managed in the same
way (new request or new rerun), C (cancellation of
a request) can be managed if the test is not in
progress. P and Q are not managed.
Relevant Clinical Information
Date and Time
Specimen Received
2- ASTM Compatible Version
2-4- The Order Block (continued)
IDS-iSYS Connection Protocol - Revision N1"
2- ASTM Compatible Version
" 10
Software version V 14