EasyManuals Logo

IDS IDS-iSYS User Manual

IDS IDS-iSYS
263 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #122 background image
Message
Meaning
CV tolerance has failed on
at least one of the
calibrators
Immunoassay
The CV calculated for one of the calibrators replicates is higher than the
percentage specified in the analytical configuration.
The CVM message is displayed with the corresponding calibrator in the
tab.
Control out of range
All fields
The control carried out with the calibration has not been calculated or is
outside the limits of acceptability.
Determination coecient
too low for one assay
Biochemistry
The slope of the kinetics calculated for one of the calibrators was calculated
on a non-linear section of the curve.
No control for checking
All fields
The control measurements requested with the calibration have not been
performed. Calibration is not considered to be valid.
Out of the activity limits
Biochemistry
The slope of the kinetics calculated for one of the calibrators is higher than
the limit value of activity specified in the analytical configuration, or enzymatic
hyperactivity was detected.
Messages
4-1- Messages Associated with Calibrations (continued)
IDS-iSYS User Manual - Revision M1!
Messages 4-1
106
Software version V 14

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the IDS IDS-iSYS and is the answer not in the manual?

IDS IDS-iSYS Specifications

General IconGeneral
BrandIDS
ModelIDS-iSYS
CategoryMeasuring Instruments
LanguageEnglish