EasyManuals Logo

IDS IDS-iSYS User Manual

IDS IDS-iSYS
263 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #239 background imageLoading...
Page #239 background image
O|1|01^7148345||^^^25OHD|R||||||A||||S||||||||||O
This block represents the information about one or several test requests for a patient.
Here are the various fields of this block used in the protocol version:
* The field value corresponds to the above block
ASTM Name of the field
Value*
Description and use in IDS-iSYS system
Record Type ID
O
Fixed value O for indicating an Order block.
Sequence Number
1
Number of the order within the Patient block starts
at 1 and is increased for each Order block in the
message.
Container and Specimen ID
01^7148345
This field consists of 2 components:
The type of container used for the sample
(2 digits).
01 PT_11_4.5 mL.
02 PT_13_6 mL.
03 ST_13_5 mL H.
05 Cup 500 µL.
06 Cup 2 mL.
07 PT_16_10 ml.
08 to 17 for configurable containers.
PT for Primary Tube; ST for Secondary Tube
For correct functioning of the apparatus, it is
important to allocate the correct type of container
used for the sample.
If this component is null, the default container type
defined for the tray will be used.
The SID of the profile concerned by the request.
Instrument Specimen ID
Not used.
Universal Test ID
^^^25OHD
This ASTM field consists of 5 components:
Universal test ID.
Universal test ID Name.
Universal test ID Type.
Manufacturer’s Code.
Sample type (patient or control)
In the direction LIS => IDS-ISYS, only the
Manufacturers code is used and this last
component will be entered with the unique IDS-
iSYS identifier of the analyte.
In one single order it is possible to specify several
tests by adding a repeat delimiter and by then
adding one or several UTI thus obtaining a chain of
the type ^^^25OHD\^^^BAP.
Multiple requests of the same test can also be made:
^^^25OHD\^^^25OHD
In the direction IDS-ISYS => LIS, the last
component will be NULL for a patient or equal to
02 for a control.
4- ASTM Compatible V3 Version
4-4- The Order Block
IDS-iSYS Connection Protocol - Revision N1"
4- ASTM Compatible V3 Version