The connection with the LIS is initialised when opening the IDS-iSYS application.
Profiles can be sent by the LIS by downloading of the work list and/or in answer to a request sent by IDS-
iSYS (the two modes are managed interchangeably).
When a new SID is detected on IDS-iSYS (barcode reading or keyboard entry), IDS-iSYS examines the work
list first of all.
If the associated profile is not in the work list, IDS-iSYS sends a request to the LIS. The LIS must send a
message in answer to each request.
If the LIS does not answer to a request, IDS-iSYS will not send or accept a new message.
When no profile is associated to a request sent by IDS-iSYS, the LIS must send a message containing a Test
Order block with the Report Type field fixed at the value Z.
In answer to a request, the LIS must send a message containing a Test Order block with the profile and with
the Report Type field imperatively fixed at the value Q.
The «Specimen type» in the field «Specimen Descriptor» (see Order Block, page 9) must be sent with the
value corresponding to the sample type: S for Serum, U for Urine (in this case, verify that the assay protocol
is defined for this sample type). The type «Other» (value =O) must not be used for Immunoassays.
In the case of a field containing multiple components, if the last component(s) takes a NULL value, the last
component delimiter(s) can be not sent.
The exchanges between IDS-iSYS and LIS can be recorded by selecting the option available in the tab
under SET-UP/SETTINGS/SYSTEM menu:
The transfers between both systems will be recorded in a log file, in the interface software folder (folder GUI).
The log must be activated only for a sort period of time, to solve transfer problems.
Activating the log permanently may cause interface software dysfunctions.
The barcode should contain at least 5 digits for the removable rack configuration (sample tubes identified by
the barcode reader integrated into the sample compartment) or at least 4 digits (except if Interleaved 2/5) for
the fixed tray configuration (sample tubes identified by the barcode reader in the front of the analyser).
If data sent by the LIS cannot be processed, the message is accepted but the corresponding profile will not
be created in the worklist.
For system equipped with the option «Clot detection», when a clot is detected in a sample, all the results
previously transferred to the LIS will be resent with a field Report Type set at the value C (correction) in the
Test Order block.
In the Result block, the field Result abnormal Flags is then forced to A.
With the versions ASTM Compatible V2 and ASTM Compatible V3, a message FIB is sent in the Comment
block.
7- Important notes
IDS-iSYS Connection Protocol - Revision N1"
7- Important notes
" 67
Software version V 14