Open the lid and insert the lid support tool (see pages 35 and 36). The sampling arm stays at its position.
•
Install the specific protective tray under the probe while holding the sampling head.
•
Remove the clot under the probe.
•
Then clean the probe with an absorbent paper soaked in alcohol at 70% v/v (ethyl or isopropyl).
•
Remove the protective tray.
•
Remove the lid support tool and close the lid (see page 36), then click on . The cycle
resumes at the end of the rinse.
Probe blockage alarm
If the probe becomes clogged when dispensing into a cuvette or when rinsing, the arm stops at the position
where the fault is detected.
•
Click on .
A message confirming the request displays:
•
To stop the cycle, click on .
If assays are under way, the analyser finishes these before stopping.
•
Then, select the fault and click on .
The following message displays:
•
Open the lid and insert the lid support tool (see pages 35 and 36).
•
Install the specific protective tray under the probe while holding the sampling head.
•
Unscrew the probe.
•
Unclog the probe (see Section 5, 5-4-11- Unclogging the sample probe, page 141).
•
Then clean the probe with an absorbent paper soaked in alcohol at 70% v/v (ethyl or isopropyl).
•
Click on . The following message displays: