Open the lid and insert the lid support tool (see pages 35 and 36). The sampling arm stays at its position.
Install the specific protective tray under the probe while holding the sampling head.
Remove the clot under the probe.
Then clean the probe with an absorbent paper soaked in alcohol at 70% v/v (ethyl or isopropyl).
Remove the protective tray.
Remove the lid support tool and close the lid (see page 36), then click on . The cycle
resumes at the end of the rinse.
Probe blockage alarm
If the probe becomes clogged when dispensing into a cuvette or when rinsing, the arm stops at the position
where the fault is detected.
Click on .
A message confirming the request displays:
To stop the cycle, click on .
If assays are under way, the analyser finishes these before stopping.
Then, select the fault and click on .
The following message displays:
Open the lid and insert the lid support tool (see pages 35 and 36).
Install the specific protective tray under the probe while holding the sampling head.
Unscrew the probe.
Unclog the probe (see Section 5, 5-4-11- Unclogging the sample probe, page 141).
Then clean the probe with an absorbent paper soaked in alcohol at 70% v/v (ethyl or isopropyl).
Click on . The following message displays:
Use
3-8- Assays (continued)
IDS-iSYS User Manual - Revision M1!
Use 3-8
74
Software version V 14