7–28
Transpector CPM Operating Manual
7 Install the ceramic shield removed in step 6 onto the new Electron Multiplier
assembly.
8 Install the new Electron Multiplier assembly into the sensor can.
9 When installing the sensor back into the feedthrough, ensure that the two pins
on the bottom of the detector align with the two pins on the feedthrough. (See
Figure 7-28.)
Figure 7-28 Pin location
10 Reverse steps 1-4 to reassemble the Transpector CPM sensor.