8–8
Transpector CPM Operating Manual
Poor 
sensitivity 
Sensor contaminated Degas sensor
Bake-out sensor
Replace ion source
Ion source pressure too low Increase process pressure, 
if possible
Verify orifice is not blocked 
and replace orifice if 
necessary
Mass calibration 
(resolution)
Adjust Mass Calibration, 
increase peak width
Sensor operating 
parameters set wrong
Check settings of: electron 
energy, ion energy, focus 
emission current
Improper calibration Ensure that the total 
pressure gauge used for 
sensitivity calibration was 
properly calibrated
Electronics failure Return to INFICON for 
repair
EM has low gain Replace sensor or EM
Poor peak shape Sensor contaminated Degas sensor
Bake-out sensor
Service sensor
Ion source pressure too 
high
Verify pressure is less than 
5E-4 Torr at the ion source
Mass calibration required Perform mass calibration
Electronics failure Return to INFICON for 
repair
Table 8-1  Symptom—Cause—Remedy Chart (continued)
SYMPTOM CAUSE REMEDY