Multitron - Incubation Shaker
Rectifying Faults
Page 110 of 146 25 February 2019
Temperature too high/low
The alarm is triggered after 20 minutes, if the deviation in the range
from 18 °C to 40 °C exceeds plus/minus 1 °C of the setpoint.
In case of temperatures outside of the above range, the alarm in
case of the same deviation is only triggered after 75 minutes.
On the example on the left, the display shows that a setpoint is ex-
ceeded (top display field).
Rotation speed too high/low
The alarm is triggered after 2 minutes, if the deviation exceeds
plus/minus 10 min
-1
of the setpoint.
On the example on the left, a shortfall of the setpoint is displayed
(bottom display field).
CO
2
concentration too high/low
The alarm is triggered after 15 minutes, if the deviation exceeds
the setpoint by more than plus/minus 0.5 %.
On the example on the left, a shortfall of the setpoint is displayed
(top display field).
Humidity too high/low
The alarm is triggered after 15 minutes, if the deviation exceeds
the setpoint by more than plus/minus 5 %.
On the example on the left, a shortfall of the setpoint is displayed
(bottom display field).
If the setpoint for a parameter is limited in the settings menu of
the equipment and an attempt is made to enter higher value, the