EasyManua.ls Logo

Infors Multitron - Specifications of the Options; Temperature Parameter (Heating and Ventilation); Internal Cooling

Infors Multitron
149 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Multitron - Incubation Shaker
Technical Data and Specifications
12.3.2 Temperature Parameter (Heating and Ventilation)
Description
Value
Unit
Power heating
800
W
Setting range
4.0 to 65.0
°C
Increment
0.1
°C
Control precision 4 50 °C
± 0.3
°C
Control precision > 50 °C
± 0.5
°C
Sensor type PT100 class 1/3 DIN B
Measurement range of the sensor
0 to 80.0
°C
Ventilation
4 axial fans, 24 VDC
Air circulation
450
m
3
/h
12.4 Specifications of the Options
The following section summarises all important technical data and
specification relating to the options.
12.4.1 Internal Cooling
Cooling unit 900 W
Electrical data
Description
Type 230 V
Type 115 V
Unit
Value
Max. power consumption
1
540 540 W
Max. current consumption 4.2 8.4 A
Fuses
(5 x 20 mm, slow-blown)
10
-
A
Thermal protection switch
-
15
A
1
) Power consumption of the compressor
Various
Description Value Unit
Cooling power
2
900
W
Additional mass
32
kg
Additional height
290
mm
2
) Cooling power at coolant temperature of 20 °C (room temperature of 20 °C)