7.2.4 Notices on Loading the Tray
Observe the following notices when loading the tray:
Place the cultivation flasks symmetrically in the middle of the
tray.
If any of the vessels protrude over the edge of the tray, they
could collide with the inner wall, resulting in the equipment be-
ing damaged or vessels breaking. As such, always ensure
that the cultivation vessels do not protrude when positioning
them on the tray.
If the tray is only loaded lightly and is going to be run at high
speeds, place additional cultivation vessels filled with water on
the tray. This will ensure that the equipment runs more
smoothly. At shaker speeds of approx. 200 up to 250 min
-1
(depending on the throw and the position of stacked equip-
ment units), the load distribution plays only a minor role.
WARNING
If the loading mass of the table is too high or too low or the load
is distributed unevenly, high rotation speeds can cause vibra-
tions. This can result in the equipment moving uncontrollably.
Do not operate the equipment without tray and loading.
In case of strong vibrations, reduce the rotation speed and
check the loading weights or the distribution of the load.