The inbound delivery and transport to the assembly location are
performed only by Infors employees or by persons authorised by
Infors.
Nonetheless it is possible that the provider’s personnel is entrusted
with transport tasks in the context of on-site transport. In this case,
observe the following notes.
10.1 Transport
WARNING
Improper transport, the use of incorrect auxiliary equipment and
careless handling of the equipment may lead to injuries and se-
vere property damage.
When transporting the equipment, observe the following:
Prior to moving the equipment, transport fasteners (rubber
wedges) must be inserted to prevent uncontrolled movements
of the table.
Always work in pairs and use suitable auxiliary equipment
when transporting the unit.
Especially when using auxiliary tools, it is important to keep in
mind that the equipment’s centre of gravity is not in the mid-
dle.
10.2 Storage
Decontaminate, thoroughly clean and dry the equipment every
time before placing it in storage. Maintain and store sensors
produced by other manufacturers in accordance with their in-
structions.
Store the equipment and its components clean, dry and pro-
tected against dust, dirt and liquids.
Store the equipment and its components in a cool place with
low humidity but protected against frost.
Storage temperature: 10 °C to 35 °C.
Relative humidity, non-condensing: 10% - 85%.
Protect the equipment from aggressive media, direct sunlight
and vibrations.