Multitron - Incubation Shaker
Installation
Page 66 of 146 25 February 2019
To achieve a stable climate in the incubation chamber and to be
able to control the climate accurately, the following installation lo-
cations must be avoided:
Poorly ventilated recesses or in the exhaust heat air stream of
air conditioners or other sources of heat
Directly underneath, next to or in the air stream of the air con-
ditioners or other devices that generate strong air flows (e.g.
circulation fans in clean rooms)
Next to equipment with strong heat radiation, such as heaters
or autoclaves, or in their exhaust heat air stream
Next to equipment for refrigeration, such as ultra-low tempera-
ture freezers or refrigerated centrifuges, or in their exhaust
heat air stream
In order to achieve the desired temperature in the incubation
chamber, the ambient temperature directly at the equipment is
decisive. This can be considerably higher than the temperature
at other positions in the room due to waste heat from the equip-
ment or other devices with strong heat radiation in the immedi-
6.2 Requirements for the Mains Connection
To void dangers due to electrical current, the in-house mains con-
nection must meet the following requirements:
Single-phase, constant power supply
In the building, the power supply must be protected using a
fault current protection switch (RCD – Residual Current De-
vice).
In addition to that, the following requirements must be met:
Make sure that the voltage values of the equipment match
those of the local power supply. Note the description on the
identification plate.
Only use the power cable provided. If the power cable be-
comes defective, replace it with a power cable of the same
type.
Ensure that the mains connection is accessible at all times.