Multitron - Incubation Shaker
Rectifying Faults
25 February 2019 Page 111 of 146
8.1.2 Alarm Message RESTARTED
The alarm message RESTARTED appears if the equipment has
switched itself back on based on the stored parameters after a
power failure.
After power is restored, the equipment automatically restarts with
the same parameters to avoid damage to the cultures due to an
unintentional termination of the process. The alarm message indi-
cates that the process has been interrupted. However, it is not pos-
sible to determine how long the interruption lasted.
The alarm message can be confirmed by tapping any key.
8.2 Faults and Error Messages
8.2.1 Error Messages Explained
The faults listed here, the causes of which are described using er-
ror messages, can generally not be resolved by users. One of the
manufacturer’s service technicians needs to be consulted if these
messages appear.
In addition to the warning indicator and the ERROR sign, mes-
sages that are abbreviated appear at the top and bottom alphanu-
meric displays. These messages have the following meanings:
The equipment door is open. When the door is opened, the
equipment is automatically stopped using the greatest possible
breaking effect.
In conjunction with the temperature parameter
(Temp symbol lights up)
Error 1, Temperature High
A temperature above 100 °C is measured.
Error 1, Temperature Low:
A temperature below 0 °C is measured.