Multitron - Incubation Shaker
Technical Data and Specifications
12.2.9 Operating and Auxiliary Materials
ATTENTION
Using the wrong auxiliary materials can result in significant dam-
age to property.
Only use the auxiliary materials in accordance with the table be-
low.
Description
Permitted/used Products
Refrigerant (cooling compres-
sor)
R134a
Coolant (cooling circuit)
30 % Antifrogen L, 70 % Water
General detergent
Mild neutral cleaning agent
Detergent for adhesive matting
Mild neutral cleaning agent
Dishwashing detergent
Disinfectant
Quaternary ammonium com-
pounds
12.3 Specifications of Standard Parameters
12.3.1 Rotation Speed Parameter (Shaker Drive)
Description
Value
Unit
Drive
BLDC motor 48 VDC
Sensor
Hall sensor
Controller
PID controller
Throw (fix throw)
3, 25 or 50
mm
Throw (adjustable throw)
12.5, 19, 25, 50
mm
RPM range
1
25 mm throw
20 to 400
min
-1
50 mm throw
20 to 350
min
-1
3 mm throw
20 to 1000
min
-1
Increment
1
min
-1
Control precision
± 4
min
-1
1
) The actual rotation speed that can be achieved depends on different fac-
tors, such as loading, vessel type (e.g. flasks with baffles) or holder (e.g.
clamps or «Sticky Stuf)