Technical Data and Specifications
25 February 2019 Page 133 of 146
Recommended rotation speeds
INFORMATION
The following information is based on an ideal tray load made
up of standard Erlenmeyer flasks with no baffles, and a fill level
of 30 %.
The following guidelines for the maximum rotation speeds are
specified to avoid damage. As such, they must be observed,
though other restrictions (e.g. tray with «Sticky Stuff» adhesive
matting) must also be taken into account.
Individual unit
3 mm
25 mm
50 mm
1000 min
-1
400 min
-1
350 min
-1
Two units stacked
3 mm
25 mm
50 mm
Upper unit
1000 min
-1
400 min
-1
350 min
-1
Lower unit
1000 min
-1
400 min
-1
350 min
-1
Three units stacked
3 mm
25 mm
50 mm
Upper unit
1000 min
-1
400 min
-1
300 min
-1
Middle unit
1000 min
-1
400 min
-1
350 min
-1
Lower unit
1000 min
-1
400 min
-1
350 min
-1
All this information is intended solely as a guideline (not guaran-
teed). Depending on the load, higher rotation speeds are possible;
in this case, the speed should be increased slowly. In such cases,
the user is responsible for determining the maximum possible
speed based on the vibrations and the vessel holders (the clamps
may need to be secured) and communicating them. If vibrations
occur, reduce the rotation speed until the equipment runs
smoothly. Alternatively, increase or reduce the rotation speed until
the equipment runs smoothly.
Maximum rotation speeds with «Sticky Stuff» adhesive mat-
ting
For a detailed description of the maximum permissible rotation
speeds when using the «Sticky Stuff» adhesive matting, see chap-
ter 5.1.3 "Tray with «Sticky Stuff» Adhesive Matting", page 55.