Multitron - Incubation Shaker
Maintenance and Cleaning
25 February 2019 Page 123 of 146
9.2.2 Disinfection
Only use quaternary ammonium compounds for wipe-down disin-
fection. As a tried as tested disinfectant, we recommend Fermaci-
dal D2.
ATTENTION
Heat (temperatures above 80 °C), aggressive disinfectants such
as chlorine bleach and UVC radiation can damage the equip-
ment and significantly limit the function and useful life of the ma-
chine.
We recommend against using UV lamps for disinfection of the
equipment because the UV rays can massively damage the
plastic casing in case of multiple applications.
For information about the procedure for disinfection underneath the
table, in particular removing and mounting the table, see chapter
9.2.1 "Cleaning", page 120.