Multitron - Incubation Shaker
Rectifying Faults
8.3 Replace Equipment Fuses
INFORMATION
The equipment fuses may only be replaced by fuses of the
same rating. For information concerning the requirements for
the fuses see chapter 12.2 "Specifications of the Basic Unit",
page 129.
To replace a defective equipment fuse, proceed as follows:
Procedure 1. Turn off the equipment and pull out the mains plug.
2. Unlock the plug for the fuses by pressing together the two
flaps and pull out the plug at the same time.
3. Remove the defective fuse.
4. Insert a new fuse with the correct number of Amperes.
5. Push the plug as far back in the opening as possible until it
snaps in.
6. Re-establish the power supply to the equipment.
8.4 Returning for Repair
The provider must return the equipment or the faulty component
part(s) to the manufacturer if, after consulting the service depart-
ment of the local dealer or the manufacturer, on-site diagnosis
and/or repair is not possible.
INFORMATION
When returning the equipment, the component part or acces-
sory for repair, it is required for the safety of all parties involved
and because of legal provisions that a lawful declaration of de-
contamination is present. Refer to main chapter “Safety and Re-
sponsibility”, chapter “Declaration of Contamination” for details.