Multitron - Incubation Shaker
Rectifying Faults
25 February 2019 Page 117 of 146
Faults in conjunction with the CO
2
parameter
CO
2
concentration too low (deviation > 1%).
2
2
Inlet pressure or flow too low.
Increase inlet pressure or flow.
2
Contact your local INFORS HT branch.
technician
or specialist dealer
CO
2
concentration too high (deviation < 1%).
CO
2
input pressure too high. Reduce input pressure. Operator
2
value fluctuates, actual value is not constant.
Inlet pressure or flow too high.
Decrease inlet pressure or flow. Operator
Faults in conjunction with the humidity parameter
The water supply for humidification is interrupted.
The water container is empty.
Top up the water container.
Air trapped between wet filter
and reservoir.
Replace the wet filter with a dry one.
Valve defective. Contact your local INFORS HT branch.
INFORS HT service
technician
or specialist dealer
The humidity parameter was deactivated automatically.
Transfer error of the humidity
sensor or temperature sensor.
Restart software (switch the equipment off and
then on again).
Inlet pressure is too high. Reduce inlet pressure (max. 0.3 bar). Operator
Steam generator is defective.
Contact your local INFORS HT branch.
technician
or specialist dealer