Multitron - Incubation Shaker
Rectifying Faults
8.2.2 Fault Tables
The following tables describe possible faults for which usually no
error message appears on the display and for which no alarm is
sounded, a few exceptions aside. In many cases operators can
resolve these themselves.
General faults
Possible cause
Measures
Personnel
Power supply of the equip-
ment is interrupted.
Check if the plugs are plugged in correctly.
Check the mains connection.
Operator
The fuse of the equipment
(230 V version) is defective or
the thermal protection switch
(115 V version) was triggered.
Change the fuse respectively reset the thermal
protection switch (see chapter 8.3 "Replace
Equipment Fuses", page 118). If the fault occurs
more than once, contact your local INFORS HT
branch.
Operator
Possible cause
Measures
Personnel
The tray is blocked by foreign
objects.
Remove foreign objects from the table.
Operator
The tray is bent. Replace the tray. Operator
Possible cause Measures Personnel
Interior lighting is deactivated.
Activate interior lighting (see chapter 7.7.7 "Set-
ting the Interior Lighting", page 106).
Operator
The LED unit of the interior
lighting is defective.
Contact your local INFORS HT branch.
INFORS HT service
technician
or specialist dealer