Rectifying Faults
Page 114 of 146 25 February 2019
8.2.2 Fault Tables
The following tables describe possible faults for which usually no
error message appears on the display and for which no alarm is
sounded, a few exceptions aside. In many cases operators can
resolve these themselves.
General faults
After activating the main switch, the display and the main switch do not light up.
Power supply of the equip-
ment is interrupted.
Check if the plugs are plugged in correctly.
Check the mains connection.
The fuse of the equipment
(230 V version) is defective or
the thermal protection switch
(115 V version) was triggered.
Change the fuse respectively reset the thermal
protection switch (see chapter 8.3 "Replace
Equipment Fuses", page 118). If the fault occurs
more than once, contact your local INFORS HT
branch.
The tray cannot be locked.
The tray is blocked by foreign
objects.
Remove foreign objects from the table.
The tray is bent. Replace the tray. Operator
The interior lighting does not work.
Possible cause Measures Personnel
Interior lighting is deactivated.
Activate interior lighting (see chapter 7.7.7 "Set-
ting the Interior Lighting", page 106).
Operator
The LED unit of the interior
lighting is defective.
Contact your local INFORS HT branch.
technician
or specialist dealer