Multitron - Incubation Shaker
Maintenance and Cleaning
Procedure 1. Open the equipment door all the way.
2. Remove the tray.
3. Turn off the main switch and pull out the plug.
4. Loosen the three screws (hexalobular M6x12) in the centre of
table using the screw driver provided.
Make sure you do not lose the screws. Otherwise, replace
them only with screws of the same length.
5. Tilt the table up by a maximum of 30°.
6. Hold the table in its tilted up position (or, ideally, get a second
person to do so), or use a suitable object to hold it there.
7. Clean the base tray with a mild detergent.
8. Thoroughly dry the floor pan using paper towels.
9. Carefully place the table on its rotary axis. When doing so,
align the rotary axis so that the threaded holes on the table
are positioned directly over the holes on the rotary axis.
10. Insert the screws and tighten them (tightening torque: 12 Nm).
INFORMATION
If any of the screws are lost or damaged, make sure that they
are only replaced with original screws (M6x12).