Multitron - Incubation Shaker
Maintenance and Cleaning
25 February 2019 Page 121 of 146
Splashing water
Always use a wet cloth when cleaning the base tray; never pour
water into the tray. Make sure that no water splashes into the bear-
ings or the electronics behind the cover plate.
After cleaning the equipment, especially the interior and the base
tray, dry it with a cloth.
ATTENTION
Water that splashes into the bearings and electronics behind the
sheet cover can damage the equipment.
As such, cleaning must be performed carefully and using as little
liquid as possible.
Dust
Over time, dust can collect in the vents and on other exposed parts
(e.g. heat exchangers). It can be removed carefully using a vac-
uum cleaner.
Installing/dismantling the table
In order to allow cleaning of the base tray in case of glass break-
ages or soiling caused by culture liquids, the table can be detached
from the counterweight and folded up. To do so, proceed as fol-
lows:
CAUTION
After being folded up, the table must be held in position by hand.
If the table is not secured properly, there is a risk of it folding
back down again. This can cause injuries.
If possible, have a second person hold the table in place
once it has been folded up. Alternatively, you can place an
object under the table to stop it from folding back down.
Proceed with caution when folding the table back down.